- Test items:pH,pCO2, pO2 ,Hct
- Sample:whole blood, serum, plasma, dialysate, CSF
- Analysis time:<90s
- Sample volume: 95μL Whole blood;50μL Capillary blood
Blood Gas Analyzer, BGA-95-4
Description
Description
This new designed blood gas analyzer has a convenient disposable calibrants cartridge containing flush solution, mixed gas and liquid calibrants, waste container. This new cartridge design can avoid operator biological hazards and environment pollution.
Features
- Intuitive touch screen
High resolution 10.4’’ touch screen
User-friendly operation interface
Parameter area provides analyzer status at a glance
Online operation and self-diagnostic tutorials
- Infrared human-detector
- Flexible and simple calibrants cartridge
High and low volume testing cartridge can be chose(300,200,150,100,50).
30 days available on board
Stored at room temperature
All-in-one design without extra consumables
Cartridge expiration alarm and residue information
- Auto QC cartridge
Optional Auto QC cartridge containing 3 levels QC design
Auto H/M/L QC once per days, customer can define
30 days available on board
Stored at room temperature
All-in-one design without extra consumables
Cartridge expiration alarm and residue information
Auto-stop analysis sample once failing on QC
- Reliable and convenient measurement unit
Capillary Sample volume only 50μL
Whole blood volume only 95μL
Multi-combination parameters
Self-made long life maintenance free electrodes
Maximum 90s result from aspiration, less than 120s from sample to sample
Long-lasting fluid system
Preheats reagents and sample
- Easy to use
The sample adaptor is suitable for both syringe and capillary samples.
The sample probe’s wiper provides convenience, sample integrity and user safety
- Fast data input
Easy sample and patient ID input
Convenient bar-code calibrants and Auto-QC cartridge replace
Easy bar-code consumables identification
- Rich data management
RS-232 interface
LIS connection under HL7 protocol
Data backup with USB port
- Compact, flexible
Size(mm) W*H*D: 400×574×344
Easy move with rolling stand
Nickel-metal hydride battery,3.0Ah
- Test Menu
pH, pCO2 , pO2 , Hct
Specifications
Measured Parameters
| Analyte | Measuring range | Unit |
| Hydrogen ion activity(pH) | 6.000―9.000 | pH scale |
| Carbon dioxide | 0.667―26.67 | kPa |
| partial pressure(pCO2) | 5.0―200.0 | mmHg |
| Oxygen partial pressure(pCO2) | 0―106.7 | kPa |
| 0―800.0 | mmHg | |
| Hct | 12.0―65.0 | % |
| B.P. | 500―800 | mmHg |
Derived(Calculated) Parameters:
| pH(TC) | pH temperature corrected |
| pCO2 (TC) | pCO2 temperature corrected |
| pO2 (TC) | pO2 temperature corrected |
| TCO2 | 0-50mmol/L |
| HCO3 | 0-50mmol/L |
| Beb | -25.0-25mmol/L |
| Beecf | -25.0-25mmol/L |
| SBC | 0-50mmol/L |
| SO2% | 40-100.0% |
| AG | Anion gap |
| RI | 0.0-70.0 |
| AaDO2 | 0-700mmHg |
| O2SAT | Oxygen Saturation |
| tHb(c) | Hemoglobin(Calculated) |
| Sample | Whole blood, Serum, Plasma, Dialysate, CSF |
| Test items | pH,pCO2, pO2 ,Hct |
| Input parameter | Patient temperature
Hemoglobin FiO B.P. |
| Interface | Serial Line RS-232
TCP/IP interface USB port Integrated Bar-code reader |
| Analysis time | <90s |
| Sample volume | 95μL Whole blood
50μL Capillary blood |
| Calibration | Automatic, programmable1-point and 2-point calibration |
| Data storage | >10000 |
| Display | 10.4” TFT touch screen |
| Battery | Nickel-metal hydride battery 3.0Ah |
| Printer | Build-in Thermal printer70mm |
| Barcode reader | Up to 14 kinds barcode type |
| Instrument | 400(W)*574(H)*344(D)mm3 |
| CAL PAK | 258(W)*180(H)*70(D)mm3 |
| AUTOQC | 103(W)*180(H)*70(D)mm3 |
| Weight | 30Kg |
| Work environment | 15℃~30℃ |
| Storage environment | 5℃~40℃ |
| Electricity | 100-240V~50/60Hz |

Get Social