- Constant speed 200 tests/ hour
- 80 reagent positions, 40 sample positions
- 44 reaction cuvettes
Fully Automatic Biochemistry Analyzer, BA-A-280
Description
DESCRIPTION
DESCRIPTION
200 Tests/Hour, Double Reagents

Multifunctional precision sampling probe
- External & internal mirror polish, external & internal probe washing
- Dedicated sampling probe equipped with sensitive liquid sensor, timely feedback of reagent & sample residuals
- Collision protection, automatic probe depth adjusting

Excellent mixer design
- Teflon coating mixer. No water dropping(reducing carry-over)
- Excellent mixing effect with standard mixing procedure

Reagent tray & Sample tray;
- 80 reagent positions,40 sample positions
- Non-stop cooling system with peltier pad inside, 24 hours 2℃-14℃
- Bar code reader(optional)

Reaction tray
- 44 reaction cuvettes
- Reaction volume as low as 150μl
- Stable & accurate temperature (37±0.1℃)for reaction

Washing systerm
- 3*4 steps automatic washing
- Efficient cleaning to ensure accurate results and reliable diagnnosis

Advanced operating system
- User friendly software interface simple, easy-to-operation
- Sample automatic dilution(decrease, increase, normal)
- Reaction result automatic checking and warning, calibaration result automatic checking, warning alarm log
- Linear & nonlinear calibration, Multi-points calibration up to 8 points
- Quality control apply to westgard, L-T plot, cumulative
SPECIFICATIONS
SPECIFICATIONS
| General information | |
| Machine type | Random access,open reagent & locked reagent (optional) |
| Test speed | Constant 200 tests/hour (Mono reagent or double reagent) |
| Test principle | Colorimetric method,turbidimetry |
| Test method | 1 point end,2 point end,Fixed-Time,Kinetics |
| Calibration type | Linear & nonlinear |
| Sample unit | |
| Sample tray | 40 sample positions |
| Sample tube | Micro cup & test tube & blood collection tube can be used.(φ12 -13 )mm * (25 -100 )mm |
| Sample volume | 2-30µL step by 0.1µL |
| Sample probe | Liquid surface detection,timely tracking for liquid volume.Collision avoidance |
| Sample barcode | Sample barcode reader (optional) |
| Reagent unit | |
| Reagent tray | 80 reagent positions (40 for Rl & 40 for R2), Non-stop cooling system with peltier pad inside, 24 hours 2°C-14″C, open system, could accept any reagent |
| Reagent volume | 20-300µL,step by 1µL |
| Reagent needle | Liquid surface detection,timely tracking for liquid volume.Collision avoidance. |
| Regent barcode | Regent barcode reader (optional) |
| Reaction unit | |
| Reaction tray | 44 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance. |
| Reaction volume | 150µL-330µL |
| Reaction time | 10 minutes |
| Temperature | Peltier pad incubation system, reaction temperature (37°C ±0.1 °C) |
| Mixing system | |
| Mixer | paddle-type mixing,effective reducing of carry-over |
| Optical system | |
| Light source | Halogen lamp |
| Wavelength | 340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm; 700nm(4 more options) |
| Absorbance range | 0-4.0 Abs |
| Resolution | 0.0001 Abs |
| Operating system | |
| Operation system | Windows XP, Windows 7, Windows 8, Windows 10 |
| Data processing | Automatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory , cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification, user operation privilege classification management |
| Report printing | 6 formats optional, support custom pattern |
| interface | TCP/IP network interface |
| Working conditions | |
| Power supply | AC 110/220V ± 10%,50/60 Hz 650W |
| Water consumption | 8L/hour(maximum) |
| Dimension | 870mm * 690mm * 740mm(L*W*H) |
| Weight | 80kg |
| Language | All languages |


Get Social