- Fully autoclavable manual pipette
- Single-channel Adjustable Volum
- Covering 0.1~10mL
Micropipette, Fully Autoclavable Manual Type, PIPE-H
Description
The new generation of fully autoclavable manual pipette is lightweight, precise, stable and durable. It adopts new piston structure, metal screw structure, fluorine rubber compound sealing ring and other technologies to reduce the pipetting resistance. At the same time, it can withstand high temperature and high pressure sterilization, resist strong chemical corrosion, and have a long service life. In addition, volume lock helps achieve more precise pipetting operations.
Specifications
Single-channel Adjustable Volume Pipettes
(9 models, covering 0.1μL~10mL.) |
||||||
Volume Range | Increment | Test Volume | Error limits in accordance with ISO8655-2 | |||
(Accuracy error) | (Precision error) | |||||
% | μL | % | μL | |||
0.1-2.5μL | 0.002μL | 2.5μL | 1.4% | 0.04 | 0.7% | 0.02 |
1.25μL | 2.5% | 0.03 | 1.5% | 0.02 | ||
0.25μL | 12.0% | 0.03 | 6.0% | 0.02 | ||
0.5-10μL | 0.01μL | 10μL | 1.0% | 0.10 | 0.4% | 0.04 |
5μL | 1.5% | 0.08 | 0.8% | 0.04 | ||
1μL | 2.5% | 0.03 | 1.8% | 0.02 | ||
2-20μL | 0.02μL | 20μL | 1.0% | 0.20 | 0.3% | 0.06 |
10μL | 1.2% | 0.12 | 0.6% | 0.06 | ||
2μL | 5.0% | 0.10 | 1.5% | 0.03 | ||
10-100μL | 0.1μL | 100μL | 0.8% | 0.80 | 0.2% | 0.20 |
50μL | 1.0% | 0.50 | 0.3% | 0.15 | ||
10μL | 3.0% | 0.30 | 1.0% | 0.10 | ||
20-200μL | 0.2μL | 200μL | 0.6% | 1.20 | 0.2% | 0.40 |
100μL | 1.0% | 1.00 | 0.3% | 0.30 | ||
20μL | 2.5% | 0.50 | 0.7% | 0.14 | ||
30-300μL | 2μL | 300μL | 0.6% | 1.80 | 0.2% | 0.60 |
150μL | 1.0% | 1.50 | 0.3% | 0.45 | ||
30μL | 2.5% | 0.75 | 0.7% | 0.21 | ||
100-1000μL | 1μL | 1000μL | 0.6% | 6.00 | 0.2% | 2.00 |
500μL | 1% | 5.00 | 0.2% | 1.00 | ||
100μL | 3% | 3.00 | 0.6% | 0.60 | ||
1000-5000μL | 5μL | 5000μL | 0.6% | 30.00 | 0.2% | 10.00 |
2500μL | 0.6% | 15.00 | 0.3% | 7.5 | ||
1000μL | 1.2% | 12.00 | 0.3% | 3 | ||
2000-10000μL | 10μL | 10000μL | 0.6% | 60.00 | 0.2% | 20 |
5000μL | 0.8% | 40.00 | 0.2% | 10 | ||
2000μL | 1.5% | 30.00 | 0.3% | 6 | ||
*For accurate and reliable pipette calibration pipette conforms to ISO 8655:2002 standards |