| ECG |
Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V) |
| Lead selection: I, II, III, avR, avL, avF, V, |
| Waveform: 2 ch |
| Lead mode: 3 Leads ( R, L, F or RA, LA, LL) |
| Lead selection: I, II, III, |
| Waveform: 1 ch |
| Gain: ′2.5mm/mV, ′5.0mm/mV, ′10mm/mV, ′20mm/mV, auto |
| HR and Alarm |
| Range Adult: 15 ~ 300 bpm |
| Neo/Ped: 15 ~ 350 bpm |
| Accuracy : ± 1% or ± 1bpm,which great |
| Resolution: 1 bpm |
| Sensitivity: > 200 (uV P-P) |
| Differential Input Impedance: > 5 MΩ |
| CMRR |
Monitor: > 105 dB |
| Operation: > 105 dB |
| Diagnosis: > 85 dB |
| Electrode offset potential: ±300mV |
| Leakage Current: < 10 uA |
| Baseline Recovery: < 3 S After Defi. |
| ECG Signal Range: ±8 m V ( Vp-p ) |
| Bandwidth |
| Surgery: 1 ~ 15 Hz |
| Monitor: 0.5 ~ 35 Hz |
| Diagnostic: 0.05 ~ 100 Hz |
| Calibration Signal: 1 (mV p-p), Accuracy:±5% |
| ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV |
| ARR Detecting |
| Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, |
| TACHY, BRADY, MISSED BEATS, PNP, PNC |
| Alarm: Available |
| Review: Available |
| RESP |
Method: Impedance between R-F(RA-LL) |
| Differential Input Impedance: >2.5 MΩ |
| Measuring Impedance Range: 0.3~5.0Ω |
| Base line Impedance Range: 0 – 2.5 KΩ |
| Bandwidth: 0.3 ~ 2.5 Hz |
| Resp. Rate |
Measuring and Alarm Range |
| Adult: 0 ~ 120 rpm |
| Neo/Ped: 0 ~ 150 rpm |
| Resolution: 1 rpm |
| Accuracy : ±2 rpm |
| Apean Alarm: 10 ~ 40 S |
| NIBP |
Method: Oscillometric |
| Mode: Manual, Auto, STAT |
| Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min) |
| Measuring Period in STAT Mode: 5 Min |
| Pulse Rate Range: 40 ~ 240 bpm |
| Alarm Type: SYS, DIA, MEAN |
| Measuring and alarm range |
| Adult Mode |
| SYS: 40 ~ 270 mmHg |
| DIA: 10 ~ 215 mmHg |
| MEAN: 20 ~ 235 mmHg |
| Pediatric Mode |
| SYS: 40 ~ 200 mmHg |
| DIA: 10 ~ 150 mmHg |
| MEAN:20 ~ 165 mmHg |
| Neonatal Mode |
| SYS: 40 ~ 135 mmHg |
| DIA:10 ~ 100 mmHg |
| MEAN:20 ~ 110 mmHg |
| Resolution Pressure:1mmHg |
| Accuracy Pressure Maximum Mean error: ±5mmHg |
| Maximum Standard deviation: ±8mmHg |
| Overpressure Protection |
| Adult Mode: 297±3 mmHg |
| Pediatric Mode: 240±3 mmHg |
| Neonatal Mode: 147±3 mmHg |
| SpO2 |
Measuring Range: 0 ~ 100 % |
| Alarm Range: 0 ~ 100 % |
| Resolution: 1 % |
| Accuracy : 70% ~ 100% ±2 % |
| 0% ~ 69% unspecified |
| Actualization interval: about 1 Sec. |
| Alarm Delay: 10 Sec. |
| Pulse Rate |
| Measuring and Alarm Range: 20~300bpm |
| Resolution: 1bpm |
| Accuracy : ±2bpm |
| TEMPERATURE |
| Channel: 2 |
| Measuring and Alarm Range: 0 ~ 50 °C |
| Resolution: 0.1°C |
| Accuracy : ±0.1°C |
| Average Time Constant:<10 Sec. |
| IBP |
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
| ART: 0 ~ 300 mmHg |
| PA: -6 ~ 120 mmHg |
| CVP/RAP/LAP/ICP: -10 ~ 40 mmHg |
| P1/P2: -10 ~ 300 mmHg |
| Press Sensor |
| Sensitivity: 5 uV/V/mmHg |
| Impedance: 300-3000Ω |
| Resolution: 1 mmHg |
| Accuracy : ±2% or ±1mmHg, which great |
| Actualization interval: about 1 Sec |
| Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature |
| Optional: Nellcor SpO2, EtCO2,IBP, Touch screen, Thermal Recorder,Wall mount, Trolley, Central station,VGA |
Get Social