- Classification: Class II Type A2
- Cleanliness level: ISO level 4
- Filters: ULPA(U15)*2pcs
- Vertical laminar negative pressure model
- Working zone area: 1300*520*640 mm
Class II Type A2 Biosafety Cabinets, BSC-IIA2-4J
Description
BSC series Class II biosafety cabinets are biological safety isolation equipment used in biological safety laboratories or other laboratories to protect personnel, tested samples and the environment. It can meet the operation of pathogenic body with risk level of 1, 2 and 3.
High-Quality & Ergonomic Construction
- The outer box is made of high-quality cold-rolled steel plate with electrostatic spraying;
- The three sides of the operation area are made of stainless steel integrated structure;
- Internal washable parts are treated with 10mm large rounded corners, which is easy to clean;
- 10° inclined design, making the operator more comfortable;
- The exhaust air is filtered by a high-efficiency filter to protect the environment from pollution;
Energy-efficient
- Class II A2 type biological safety cabinet;
- Vertical laminar negative pressure model, air flow mode: 30% discharge, 70% internal circulation
High security
- Negative pressure design ensures the safety of operators;
- Integrated sliding safety door positioning, easy to operate;
- Equipped with filter blocking alarm, blower overload alarm, working window opening limit alarm system;
Specifications
Model | BSC-IIA2-4J |
Classification | Class II Type A2 |
Cleanliness level | ISO level 4 |
Filter efficiency | ≥99.9995%,@0.12μm |
Filters | ULPA(U15)*2pcs |
Average airflow velocity of downdraft(m/s) | 0.25~0.40 |
Average airflow velocity of inflow air(m/s) | ≥0.5 |
Sound emission (dBA) | ≤67 |
Vibration peak (μm) | ≤5 |
Electricity | AC single-phase 220V/50Hz |
Air balance biological protection | Personnel protection: a.The total number of colonies in the impact sampler:≤10CFU/time b.Total colony count of the slot sampler:≤5CFU/time |
Product protection1-8CFU/ml(repeat 3times,5min/time)total number of colonies:≤5CFU/time | |
Cross contamination protection1-8CFU/ml(repeat 3times,5min/time)total number of colonies:≤5CFU/time | |
Max. power(KV·A) | ≤1.5 (Spare socket included) |
Net weight/Gross weight(Kg) | 270/280 |
Working zone area(W*D*H) | 1300*520*640 mm |
External dimension(W*D*H) | 1465*760*2200mm |
LED lamp | 20W*2pcs |
UV lamp | 30W*1pc |
Illumination (Lux) | 900 |
Supply and exhaust pipe diameter(mm) | Φ250 |
Wind direction | Top out |