- Classification: Class II Type B2
- Cleanliness level: ISO level 4
- Filters: ULPA(U15)*2pcs
- Air flow mode: 100% discharge, 0 internal circulation
- Vertical laminar negative pressure model
- Working zone size: 1600*520*640mm
Class II Type B2 Biosafety Cabinets, BSC-IIB2-5J
Description
BSC biosafety cabinets are biological safety isolation equipment used in biological safety laboratories or other laboratories to protect personnel, tested samples and the environment. It can meet the operation of pathogenic body with risk level of 1, 2 and 3.
High-Quality & Ergonomic Construction
- The outer box is made of high-quality cold-rolled steel plate with electrostatic spraying;
- The three sides of the operation area are made of stainless steel integrated structure;
- Internal washable parts are treated with 10mm large rounded corners, which is easy to clean;
- 10° inclined design, making the operator more comfortable;
- The exhaust air is filtered by a high-efficiency filter to protect the environment from pollution;
Energy-efficient
- Class II type B2 biological safety cabinet;
- Vertical laminar negative pressure model, air flow mode: 100% discharge, 0 internal circulation.
High security
- Negative pressure design ensures the safety of operators;
- Integrated sliding safety door positioning, easy to operate;
- Equipped with filter blocking alarm, blower overload alarm, working window opening limit alarm system;
- The exhaust air is filtered by a high-efficiency filter to protect the environment from pollution;
Energy-efficient
- Class II type B2 biological safety cabinet;
- Vertical laminar negative pressure model, return air 30%/100% filtration discharge;
High security
- Negative pressure design ensures the safety of operators;
- Integrated sliding safety door positioning, easy to operate;
- Equipped with filter blocking alarm, blower overload alarm, working window opening limit alarm system;
Specifications
Model | BSC-IIB2-5J |
Classification | Class II Type B2 |
Cleanliness level | ISO level 4 |
Filter efficiency | ≥99.9995%,@0.12μm |
Filters | ULPA(U15)*2pcs |
Average flow rate of downdraft(m/s) | 0.25~0.40 |
Average flow rate of inflow air(m/s) | ≥0.5 |
Sound emission (dBA) | ≤67 |
Vibration peak(μm) | ≤5 |
Electricity | AC single-phase 220V/50Hz |
Air balance biological protection | Personnel protection: a.The total number of colonies in the impact sampler:≤10CFU/time b.Total colony count of the slot sampler:≤5CFU/time |
Product protection1-8CFU/ml(repeat 3times,5min/time) total number of colonies:≤5CFU/time |
|
Cross contamination protection 1-8CFU/ml(repeat 3times,5min/time) total number of colonies:≤5CFU/time | |
Max power(KV·A) | ≤1.5(Spare socket included) |
Weight(Kg) | 300/310 |
Working zone area W*D*H(mm) |
1600*520*640 |
External dimension W*D*H(mm) |
1765*760*2200 |
LED lamp | 25W*2pcs |
UV lamp | 40W*1pc |
Illumination (Lux) | 900 |
Supply and exhaust pipe diameter(mm) | Φ250 |
Wind direction | Top in,top out |