- 2 to 5 points calibration
- Unit switches: NTU, FNU, EBC and ASBC
- Suitable for process control and field use
- Accuracy: ±2% of reading (0~500NTU), ±3% of reading (501~1100NTU)
Portable Turbidity Meter, BEP-TB100
Description
Feature:
- A high-performance portable turbidity meter meets the design criteria in ISO 7027
- 2 to 5 points calibration using the Formazin Standards
- Selectable 4 turbidity units, including the NTU, FNU, EBC and ASBC
- Single measurement mode automatically senses and locks a stable reading
- Continuous measurement mode can be used for indexing or matching the sample vials
- Auto-Power Off effectively conserves battery life
- The setup menu allows setting the number of calibration points, resolution, date and time, etc.
- Reset function automatically resumes all settings back to the factory defaults
- Expanded memory stores or recalls up to 100 data sets
- USB communication interface for data transfer or connecting a power adapter to meter
Ordering Information
BEP-TB100: Meter, turbidity standards (0.02, 200, 500, 1000 NTU), sample vials, lint-free cloth, and carrying case
Specifications
| Model | BEP-TB100 | |
| Turbidity | Principle | ISO 7027 nephelometric method (90°) |
| Range | 0~1100NTU, 0~275 EBC, 0~9999ASBC | |
| Resolution | 0.01 (0~99NTU), 0.1 (100~999NTU), 1 (1000~1100NTU) | |
| Accuracy | ±2% of reading (0~500NTU), ±3% of reading (501~1100NTU) | |
| Calibration Points | 2 to 5 points | |
| Calibration Standards | 0.02, 10, 200, 500, 1000NTU | |
| Other Specifications | Light Source | Infrared-emitting diode (850nm wavelength) |
| Detector | Silicon photo diode | |
| Stray Light | <0.02 NTU | |
| Sample Vial | 60(H)× 25 (Dia)mm | |
| Sample Volume | 30mL | |
| Memory | 100 data sets | |
| Communication Interface | USB | |
| Operating Temperature | 0-50℃ | |
| Display | Custom LCD (60×40mm) | |
| Power Requirements | 3×1.5V AA batteries or DC5V power adapter | |
| Dimensions | 180 (L)× 85 (W)×70 (H)mm | |
| Weight | 300g | |