- LCD display.
- Data Storage500 sets.
- Support for USB or RS-232.
- IP54 waterproof.
Benchtop pH/Ion Meter, BEP-M400B
Description
Description
The BEP-M400B Benchtop pH/Ion Meter is suitable for health and disease control, environmental protection, biological agriculture and forestry, water quality analysis, scientific research and higher institutes and other fields.The ion-selective electrode method has the advantages of simple operation and low detection limit.
Features:
General Features
- High resolution LCD display, 5.7 inches.
- Multi-reading feature allows auto-read, timed-read and continuous-read.
- Automatic/Manual temperature compensation ensures accurate results.
- Auto-hold feature senses and locks the measurement endpoint.
- Data Storage500 sets (GLP-compliant).
- Support for USB or RS-232 communication.
- Reset feature automatically resumes all settings back to factory default options.
- IP54 waterproof.
pH:
- 1-5 points calibration with Standard Recognition.
- Selectable pH buffer groups, including NIST, DIN, GB.
- Automatic electrode diagnosis with pH slope and offset display.
Ion:
- 1-5 points calibration.
- Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, PX, etc.
- Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.
- Over 10 methods are built-in, including F–, Cl–, Br–, I–, NO3–, BF4–, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+and etc., user-defined method is supported.
Specifications
| Model | BEP-M400B | |
| Parameters | pH/ISE/Temp.(mV/pX) | |
| pH | Range | -2.000 to 20.000 pH | 
| Resolution | 0.1, 0.01, 0.001pH | |
| Accuracy | ±0.002 pH | |
| Calibration Points | Up to 5 | |
| Standard Customization | Yes | |
| Standard Recognition | NIST, GB, DIN, USA and MERK buffers | |
| Slope Limit | Yes | |
| mV | Range | -2000.00 to 2000.00 mV | 
| Resolution | 0.1, 0.01 mV | |
| Accuracy | ±0.1 mV or ±0.03% of reading whichever is greater | |
| pX | Range | -2.000 to 20.000 pX | 
| Resolution | 0.1, 0.01, 0.001 pX | |
| Accuracy | ±0.002 pX | |
| Calibration Points | Up to 5 | |
| ISE | Range | 1E-9 to 9.999E9 | 
| Unit | mol/L, mmol/L, g/L, mg/L, μg/L, ppm | |
| Resolution | Up to 4 significant digits | |
| Accuracy | ±0.3% | |
| Calibration Points | Up to 5 | |
| Temperature | Range | -10 to 135 ºC, 14 to 275 ºF | 
| Unit | ºC, ºF | |
| Resolution | 0.1 | |
| Accuracy | ±0.1 | |
| Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous | 
| Reading Prompts | Reading, Stable, Locked | |
| Temp. Compensation | ATC, MTC | |
| Data management | Data Storage | 500 results each | 
| GLP Features | Yes | |
| Inputs | pH Electrode | BNC(Q9) | 
| Temp. Probe | 4-pin aviation connector | |
| Outputs | USB | USB 2.0 flash memory device, PC | 
| RS 232 | printer | |
| Display options | Backlight | Yes | 
| Auto Shutdown | 1~60 min, off | |
| IP Rating | IP54 | |
| Date and Time | Yes | |
| General | Electricity | AC Adapter, 100-240V AC input, DC9V output | 
| Dimensions | 242×195× 68 mm | |
| Weight | 900g | |
 
											
				 
			
											
				
 
	 
						 
	 
						 
						 
						
Get Social